Freeze-Dried Lyophilised PCR Mixes Description:
This kit is used for the detection of COVID-19 (ORF lab/N Gene) Nucleic Acid and assisted diagnosis and epidemiological surveillance of COVID-19.
Lyophilised PCR Mixes Composition:
COVID-19 PCR Mix、Resuspension Buffer、COVID-19 Negative Control、COVID-19 Positive Control
Package Instruction:
50 tests/kit & 100 tests/kit
Packing: Paper、Foil pouch
Storage Temperature: -20±5℃
Period of Validity: 6 Months
Freeze-Dried Lyophilised PCR Mixes Features
1. Multiplex One-step RT Real-time PCR assay for reliable test result.
2. Good performance with high sensitivity and specificity co-developed and testified
by Wuhan Institute of Virology, Chinese Academy of Sciences.
3. Compatible with varied Real-time PCR systems.
4. Sensitivity:2500 copies/mL, Reaction time: 1h.
5. Low requirements for transportation environment
6. Low transportation cost of products
Specifications and Models:
Name | Number | Packing unit |
New Coronavirus COVID-19 Freeze-Dried Lyophilised PCR Mixes | 50 or 100 tests | Box |
Weight Volume of Real Time PCR Detection Kit:
Unit | Piece | Volume(CM) | Weight(KG) |
per box | 50 | 5.5*5.5*5.5 | 0.025 |
per carton | 80*50 | 65*40*45 | 4.35 |
per pallet | 14*80*50 | 120*105*135 | 60.9 |
New Coronavirus COVID-19 Freeze-Dried Lyophilised PCR Mixes Result Judgment:
1. If the test sample detects a typical S-type amplification curve in the FAM and
HEX channels and the Ct value is <40, and there is a typical S-type amplification curve
in the Internal Control channel (Cy5), the sample can be judged to be COVID-19
positive.
2. If the test sample has no amplification curve or Ct value> 40 in the FAM and HEX
channels, and there is a typical S-type amplification curve in the Internal Control
channel (Cy5), the sample can be judged to be COVID-19 negative.
3. If the test sample only has a Ct value≤40 in a single channel of the FAM or HEX
channel, and there is a typical S-type amplification curve in the Internal Control channel
(Cy5), the results need to retest. If the results repeated are consistent, the sample can be
judged as COVID-19 positive, the results repeated are negative except for the typical S-
type amplification curve of the Internal Control channel (Cy5), which could be judged
as COVID-19 negative.
4. If no typical S-type amplification curve (No Ct value) or Ct value> 40 is detected
in the FAM, HEX, and Cy5 channels of the test sample, it means that there is a problem
with the quality of the sample or a problem with the operation. If the result is invalid,
you should find and eliminate the cause, collect the sample again, and repeat the test (if
the test result is still invalid, please contact the company).